Case Overview

Testing Period

December 28, 2023 –   March 31, 2024

Testing Location

Dongfeng Chemical Industrial Park, Taizhou   City, Zhejiang Province

Testing Environment Temperature:

Temperature: 21–24°C, Humidity: 42–50% RH

Testing Content

LaSense, based on its independently developed   cavity-enhanced CEAS technology, developed the first batch of domestically   produced high-precision greenhouse gas laser analyzers, the MT01, which were   used to monitor CO concentrations   at multiple industrial exhaust outlets in the Taizhou Industrial Park,   Zhejiang Province.


Case Results

Overall Performance

 

The equipment monitors multiple emission points simultaneously via an automatic sampling system. The instrument has been operating continuously without interruption, with zero maintenance required. The lowest CO emission measured was 40 ppm, and the highest exceeded 9,000 ppm. The instrument is calibrated monthly using standard gases, with a deviation of less than 0.1 ppm, meeting user requirements for precision, repeatability, dynamic range, and other instrument parameters.

 